Radiation safety Interlock System

Radiation safety Interlock System

Safety door interface panel(FS-SIP-01) is designed for interfacing with imaging equipment. The purpose of this panel is to prevent unwanted radiation exposure for the general public, also to make sure no interruption of scanner while a procedure is carried by locking the door. The working principle of the panel is to prevent the scanner or imaging equipment to release radiation. The panel also disables the door from opening when the scanner is emitting radiation.

STATUS OF DOOR AND RADIATION FROM AN EQUIPMENT :


Door Position

Door Open

Radiation cannot be released

Door Closed

Radiation can be released


Radiation Machine Status

Radiation Off

Door can be opened

Radiation On

Door cannot be opened


INSTALLATION


The radiation interlock system can be installed at any existing site. The installation will take only one day to complete, the door lock will be part of the radiation interlock system, the door interlock system can be also connected to BMS and access door system of the hospital.




Radiation Door Interlock Module
FME-M-1 FME-M-

http://www.frankshield.com/images/fme_m.jpg      http://www.frankshield.com/images/fme_m_0.jpg


 

Send Message